Performance of a Novel Continuous Glucose Monitoring Device in People With Diabetes

血糖性 医学 连续血糖监测 血糖自我监测 糖尿病 内科学 内分泌学
作者
Julia K. Mader,Delia Waldenmaier,Wiebke Mueller-Hoffmann,Katrin Mueller,Michael Angstmann,Gerhard Vogt,Cosima Rieger,Manuel Eichenlaub,Thomas Forst,Guido Freckmann
出处
期刊:Journal of diabetes science and technology [SAGE]
标识
DOI:10.1177/19322968241267774
摘要

Background: In this multicenter study, performance of a novel continuous glucose monitoring (CGM) system was evaluated. Methods: Adult participants with diabetes were included in the study. They each wore three sensors of the CGM system on the upper arms for up to 14 days. During four in-clinic visits, frequent comparison measurements with capillary blood glucose (BG) samples were performed. The primary endpoint was the 20/20 agreement rate (AR): the percentage of CGM readings within ±20 mg/dL (at BG values <100 mg/dL) or ±20% (at BG values ≥100 mg/dL) of the comparator. Further evaluations included mean absolute relative difference (MARD) and 20/20 AR in different BG ranges and across the wear time. Results: Data from 48 participants and 139 sensors were analyzed. During in-clinic sessions the 20/20 AR was 90.5% and the MARD was 9.2%. For BG ranges <70, 70-180, and >180 mg/dL, 20/20 AR was 94.3%, 89.0%, and 92.5%, respectively. At the beginning, middle, and end of sensor wear time, 20/20 AR was 92.8%, 91.5%, and 85.9%, respectively. The 14-day survival probability was 82.4%. Pain and bleeding after sensor insertion were within the expected range. Based on the study outcome, the use of the device is regarded as safe. Conclusions: The system showed a good performance compared to capillary BG measurements. This level of accuracy could be shown over the entire measurement range, especially in the low glycemic range, and the whole wear time of the sensors. The results of this study are supporting a non-adjunctive use of the device.
最长约 10秒,即可获得该文献文件

科研通智能强力驱动
Strongly Powered by AbleSci AI
更新
大幅提高文件上传限制,最高150M (2024-4-1)

科研通是完全免费的文献互助平台,具备全网最快的应助速度,最高的求助完成率。 对每一个文献求助,科研通都将尽心尽力,给求助人一个满意的交代。
实时播报
XSC发布了新的文献求助10
2秒前
3秒前
liu完成签到,获得积分20
3秒前
科研小白发布了新的文献求助10
3秒前
5秒前
易生关注了科研通微信公众号
5秒前
Kikisman发布了新的文献求助10
6秒前
liu发布了新的文献求助10
6秒前
cc应助Evan采纳,获得10
7秒前
miaolingcool发布了新的文献求助10
11秒前
李凌霄完成签到,获得积分10
11秒前
Kikisman完成签到,获得积分10
13秒前
nenoaowu发布了新的文献求助30
13秒前
14秒前
sereno完成签到 ,获得积分10
15秒前
善学以致用应助李凌霄采纳,获得10
17秒前
由哎发布了新的文献求助10
19秒前
干净的铅笔应助Enoch采纳,获得10
19秒前
20秒前
24秒前
24秒前
易生发布了新的文献求助10
24秒前
25秒前
今后应助勤劳菠萝采纳,获得10
27秒前
27秒前
27秒前
28秒前
Lone完成签到,获得积分10
28秒前
无花果应助长度2到采纳,获得20
28秒前
Gzero1完成签到,获得积分10
29秒前
siriuslee99发布了新的文献求助10
29秒前
30秒前
31秒前
31秒前
cc应助科西西采纳,获得10
31秒前
默默尔烟发布了新的文献求助10
32秒前
Orange应助绕地球3圈采纳,获得10
32秒前
33秒前
long发布了新的文献求助10
33秒前
小二郎应助霜序采纳,获得10
33秒前
高分求助中
Impact of Mitophagy-Related Genes on the Diagnosis and Development of Esophageal Squamous Cell Carcinoma via Single-Cell RNA-seq Analysis and Machine Learning Algorithms 2000
Evolution 1100
How to Create Beauty: De Lairesse on the Theory and Practice of Making Art 1000
Gerard de Lairesse : an artist between stage and studio 670
CLSI EP47 Evaluation of Reagent Carryover Effects on Test Results, 1st Edition 550
Assessment of Ultrasonographic Measurement of Inferior Vena Cava Collapsibility Index in The Prediction of Hypotension Associated with Tourniquet Release in Total Knee Replacement Surgeries under Spinal Anesthesia 500
T/CAB 0344-2024 重组人源化胶原蛋白内毒素去除方法 500
热门求助领域 (近24小时)
化学 医学 生物 材料科学 工程类 有机化学 生物化学 物理 内科学 纳米技术 计算机科学 化学工程 复合材料 基因 遗传学 催化作用 物理化学 免疫学 量子力学 细胞生物学
热门帖子
关注 科研通微信公众号,转发送积分 2981308
求助须知:如何正确求助?哪些是违规求助? 2642696
关于积分的说明 7131118
捐赠科研通 2276049
什么是DOI,文献DOI怎么找? 1207286
版权声明 592084
科研通“疑难数据库(出版商)”最低求助积分说明 589801