Sixty-degree nasal sectoral juxtacanalicular space goniotomy with the single-use Espaillat juxtacanalicular space slider device significantly reduced intraocular pressure and the number of required glaucoma hypotensive medications, with sustained reductions over 24 months.This study analyzed the safety and efficacy of the Espaillat juxtacanalicular space slider during minimally invasive 60-degree nasal sectoral goniotomy for mild-to-moderate open angle glaucoma (OAG).This prospective, observational case series performed at a private clinical practice and outpatient surgery center assessed the eyes of patients with mild-to-moderate OAG who underwent phacoemulsification cataract surgery with 60-degree nasal sectoral goniotomy using the Espaillat juxtacanalicular space slider. The same surgeon performed all procedures. Intraocular pressure, intraocular pressure-lowering drugs, and adverse effects were assessed over 24 months, and a regression analysis of intraocular pressure reduction was performed. Success was defined as (1) A reduction in the intraocular pressure of at least 20% for at least 12 months, compared with the baseline value, and (2) a decrease in the number of medications by at least one for at least 12 months.Among 38 eyes, 27 (72%) had moderate glaucoma, and 11 (28%) had mild glaucoma. Postoperatively, intraocular pressure was decreased by 38% at 12 months (mean 13.7±1.7 mmHg), 35% at 18 months (mean 14.4±1.8 mmHg), and 31% at 24 months (mean 15.2±1.9 mmHg). In addition, the number of topical glaucoma medications used decreased from a mean of 1.6 to 0.4 ( P <0.001). The main adverse event was temporary focal corneal edema (29 patients; 76%).Using the Espaillat juxtacanalicular space slider during invasive 60-degree nasal sectoral goniotomy for OAG yielded a significant and sustained reduction in intraocular pressure and reduced the need for glaucoma medications with minimal adverse events.