Irene Abicca,Domenico Schiano-Lomoriello,Marta Gilardi,Daniela Giannini,Valentin Dinu,Catarina P. Coutinho,Giacomo Savini
出处
期刊:Journal of Refractive Surgery
[SLACK, Inc.]
日期:2024-07-01
卷期号:40 (7)
被引量:1
标识
DOI:10.3928/1081597x-20240514-03
摘要
Purpose: To evaluate the repeatability of automatic measurements of a new anterior segment optical coherence tomographer (ANTERION; Heidelberg Engineering) and their agreement with an anterior segment optical coherence tomography device combined with Placido disc corneal topography (MS-39; CSO) in patients affected by keratoconus. Methods: Fifty-four consecutive patients were included. Three measurements were performed with the ANTERION and one with the MS-39. Repeatability was assessed by means of within-subject standard deviation, coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement was investigated with the 95% limits of agreement. The paired t -test and Wilcoxon matched-pairs test were performed to compare the measurements of the different devices. Results: Repeatability of ANTERION measurements was high, with an ICC greater than 0.98 for all parameters. Many parameters revealed a CoV of less than 1% and a CoV within 5% was obtained for astigmatism measurements. The ANTERION measured a significantly higher corneal power and the MS-39 more negative posterior keratometric values. These differences were mirrored by a moderate agreement for mean simulated keratometry and poor agreement for total corneal power and posterior keratometry. Conclusions: The ANTERION revealed high repeatability of automatic measurements and good agreement with the MS-39 for many parameters in patients affected by keratoconus, but for most parameters the two instruments cannot be considered interchangeable. [ J Refract Surg . 2024;40(7):e445–e452.]
科研通智能强力驱动
Strongly Powered by AbleSci AI