Joonas Paalasmaa,Hannu Toivonen,Markku Partinen
出处
期刊:IEEE Journal of Biomedical and Health Informatics
[Institute of Electrical and Electronics Engineers]
日期:2015-11-01
卷期号:19 (6): 1945-1952
被引量:104
标识
DOI:10.1109/jbhi.2014.2314144
摘要
We present a method for measuring beat-to-beat heart rate from ballistocardiograms acquired with force sensors. First, a model for the heartbeat shape is adaptively inferred from the signal using hierarchical clustering. Then, beat-to-beat intervals are detected by finding positions where the heartbeat shape best fits the signal. The method was validated with overnight recordings from 46 subjects in varying setups (sleep clinic, home, single bed, double bed, two sensor types). The mean beat-to-beat interval error was 13 ms and on an average 54% of the beat-to-beat intervals were detected. The method is part of a home-use e-health system for an unobtrusive sleep measurement.