We have developed a user-friendly hybrid surface temperature sensor. The uncertainties of temperature readings associated with this sensor and a thermocouple embedded in a silicon wafer are compared. The expanded uncertainties (k=2) of the hybrid temperature sensor and the embedded thermocouple are 2.11 and 2.37 K, respectively, in the temperature range between 600 and 1000 K. In the present paper, the uncertainty evaluation and the sources of uncertainty are described.