液体闪烁计数
水准点(测量)
计量系统
物理
光学
化学
放射化学
天文
大地测量学
地理
作者
Qianqian Zhou,Haoran Liu,Zhijie Yang,Weihua Zeng,Juncheng Liang
标识
DOI:10.1016/j.apradiso.2022.110315
摘要
In order to meet the demand of on-site measurement for radionuclides, a portable liquid scintillation TDCR system was developed at National Institute of Metrology (NIM), China. The system consists small size TDCR counter for the measurement of liquid scintillation sources, and digital electronics for pulse signal processing. The optical chamber adopts Teflon material with high diffuse reflection efficiency. Two independent signal processing solutions were used here for TDCR counting. One employed the on-line TDCR solution based on FPGA counting module named TDCR-DMCA, and the other adopted the off-line TDCR solution based on a stand-alone desktop digitizer of CAEN. Two solutions are applied to perform coincidence, dead-time and counting operations follow by MAC3 logic. The performance of the TDCR counting system was tested in benchmark comparison with the traditional custom-built TDCR counting system at NIM through activity measurements of 3H, 14C. Good agreement between these two systems was observed.
科研通智能强力驱动
Strongly Powered by AbleSci AI