Rufeng Bai,Yaju Liu,Zheng Li,Hongfang Jin,Qinghua Tian,Meisen Shi,Shuhua Ma
摘要
Abstract In this study, a new STRtyper-27 system, including the same Yfiler Plus loci (DYS19, DYS389I, DYS389II, DYS390, DYS391, DYS392, DYS393, DYS385a/b, DYS437, DYS438, DYS439, DYS448, DYS456, DYS458, DYS635, Y-GATA H4, DYS449, DYS460, DYS481, DYS518, DYS533, DYS570, DYS576, DYS627 and DYF387S1a/b), was established using a set of 5 fluorescent dye labels. Primers, internal size standard, allelic ladders and matrix standard set were designed and created in-house for this multiplex system. This paper describes the validation studies conducted with the STRtyper-27Y system using a 3130XL genetic analyzer for fragment length detection that included the analysis of the following parameters and aspects: sensitivity, species specificity, inhibition, haplotype concordance, precision, stutter, DNA mixtures, and stability studies with crime scene samples. The studies demonstrated, that the STRtyper-27Y system provided equivalent overall performance comparable to the latest Yfiler Plus kit, but with enhanced compatibility in terms of instrument platforms and software allowing forensic laboratories to conduct its forensic application and evaluate its performance, all in their own 5 dye Y-STR chemistry system /environment without software or instrument upgrades.