Ying Yan,Menglei Ge,Wei Wang,Haijian Zhao,Donghuan Wang,Cuihua Hu,Jing Wang,Wenxiang Chen,Chuanbao Zhang
标识
DOI:10.1016/j.cca.2016.07.004
摘要
Calcium is an important serum ion which is frequently assayed in clinical laboratories. Since quality assurance requires reference methods, the establishment of a candidate reference method for serum calcium measurement is important.An inductively coupled plasma mass spectrometry (ICP-MS) method was developed. Serum samples were spiked gravimetrically with aluminum internal standard, digested with 69% ultrapure nitric acid and diluted to test concentration. Then the (44)Ca/(27)Al ratios were measured by ICP-MS in hydrogen mode. The method was calibrated using 5% nitric acid matrix calibrators and the calibration function was established with bracketing method.The correlation coefficients between the measured (44)Ca/(27)Al ratios and the analyte concentrations ratios were all >0.9999. The coefficients of variation of the measurements were 0.27% and 0.16% for two spiked serum. The analytical recovery was 100.24%. The accuracy of the measurement was also verified through measurement of certified reference materials. Comparison with recognized reference method and international inter-laboratory comparisons gave satisfied results.New ICP-MS method is specific, precise, simple, and low in cost, and may be used as a candidate reference method in the standardization of serum calcium measurement.