Tian Zheng,Xin Yan,Fan Yang,Guidong Dai,Fang Wang,Guangxiang Chen
标识
DOI:10.1016/j.ejrad.2024.111287
摘要
To explore the optimal kiloelectron voltage (keV) of virtual monochromatic images (VMIs) of dual-layer spectral detector computed tomography (DLSCT) to display laryngeal and hypopharyngeal squamous cell carcinoma (LHSCC) and its diagnostic performance for preoperative T staging of LHSCC.A total of 67 LHSCC patients were included, and the contrast between the tumor and sternocleidomastoid muscle (SM), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and image noise of 40-100 keV VMIs and conventional polyenergetic images (CIs) were evaluated. The image quality of the CI and 40-100 keV VMI was evaluated by a five-point method. The VMI with the best image quality was screened out, and the accuracy of the optimal keV VMI and CI for T staging was assessed using clinical T staging as the reference standard.The contrast between the tumor and SM, SNR, CNR and subjective image quality scores of LHSCC on 40-50 keV VMIs were higher than those on CIs (P < 0.05); the image noises of 40-100 keV VMIs were lower than those of CIs (P < 0.05). The 40 keV VMI had the highest SNR, CNR and subjective score of image quality. The accuracy rates of the 40 keV VMI and CI for T staging of LHSCC were 0.86 and 0.63 (P < 0.001), respectively.The image quality of 40-50 keV VMI is higher than that of CI, and the diagnostic accuracy of 40 keV VMI is better than that of CI, which is most suitable for preoperative T staging of LHSCC.