Carles Igual,Jorge Igual,Janne M. Hahne,Lucas C. Parra
出处
期刊:IEEE Transactions on Neural Systems and Rehabilitation Engineering
[Institute of Electrical and Electronics Engineers]
日期:2019-01-24
卷期号:27 (2): 314-322
被引量:31
标识
DOI:10.1109/tnsre.2019.2894464
摘要
In proportional myographic control, one can control either position or velocity of movement. Here, we propose to use adaptive auto-regressive filters, so as to gradually adjust between the two. We implemented this in an adaptive system with closed-loop feedback, where both the user and the machine simultaneously attempt to follow a cursor on a 2-D arena. We tested this on 15 able-bodied and three limb-deficient participants using an eight-channel myoelectric armband. The human-machine pairs learn to perform smoother cursor movements with a larger range of motion when using the auto-regressive filters, as compared with our previous effortswithmoving-average filters. Importantly, the human-machine system converges to an approximate velocity control strategy resulting in faster and more accuratemovements with lessmuscle effort. The method is not specific tomyoelectriccontroland could be used equally well for motion control using high-dimensional signals from reinnervatedmuscles or direct brain recordings.