分子印迹聚合物
材料科学
黄曲霉毒素
检出限
傅里叶变换红外光谱
纳米技术
色谱法
化学
生物传感器
聚苯胺
生物分子
聚合物
聚合
化学工程
有机化学
选择性
复合材料
食品科学
工程类
催化作用
作者
Avinash Kumar Singh,G.B.V.S. Lakshmi,Manuel A. Fernandes,Tanushri Sarkar,Payal Gulati,Ravindra Pratap Singh,Pratima R. Solanki
标识
DOI:10.1016/j.microc.2021.106730
摘要
The sensitive electrochemical sensing platforms were fabricated using molecularly imprinted polymer (MIP) based technique for mycotoxins [Aflatoxin B1 (AFB1) and Fumonisin B1 (FuB1)] detection. Polyaniline was used as a MIP matrix synthesized using the chemical oxidative polymerization method in the presence of AFB1 and FuB1toxins as the template molecules. Fourier transform infrared spectroscopy and electrochemical techniques were used for the characterization of the synthesized MIPs. The proposed biosensors exhibited good sensitivity of 7.037 and 9.092 µA mL ng−1 cm−2 with the lower detection limits of 0.313 and 0.322 pg mL−1 for AFB1 and FuB1. Both sensors showed good linearity amidst the sensing range of 1 pg mL−1 to 500 ng mL−1. The fabricated sensors open up a promising strategy to detect AFB1 and FuB1 as an alternative for biomolecules with enormous specificity, quick responsiveness, and cost-effectiveness. These fabricated sensors manifest good analytical performance compared to other reported MIP-based sensors for AFB1 and FuB1 detection.
科研通智能强力驱动
Strongly Powered by AbleSci AI