超声波
横波
剪切(地质)
材料科学
超声波传感器
声学
传感器
声辐射力
弹性成像
高强度聚焦超声
生物医学工程
医学
物理
复合材料
作者
Ryosuke Iwasaki,Ryo Takagi,Ryo Nagaoka,Hayato Jimbo,Shin Yoshizawa,Yoshifumi Saijo,Shin-ichiro Umemura
标识
DOI:10.7567/jjap.55.07kf05
摘要
Abstract Shear wave elastography (SWE) is expected to be a noninvasive monitoring method of high-intensity focused ultrasound (HIFU) treatment. However, conventional SWE techniques encounter difficulty in inducing shear waves with adequate displacements in deep tissue. To observe tissue coagulation at the HIFU focal depth via SWE, in this study, we propose using a two-dimensional-array therapeutic transducer for not only HIFU exposure but also creating shear sources. The results show that the reconstructed shear wave velocity maps detected the coagulated regions as the area of increased propagation velocity even in deep tissue. This suggests that “HIFU-push” shear elastography is a promising solution for the purpose of coagulation monitoring in deep tissue, because push beams irradiated by the HIFU transducer can naturally reach as deep as the tissue to be coagulated by the same transducer.
科研通智能强力驱动
Strongly Powered by AbleSci AI