Lei Zhu,Jing Wang,Lei Xing
摘要
Scatter correction is crucial to the quality of reconstructed images in x‐ray cone‐beam computed tomography (CBCT). Most of existing scatter correction methods assume smooth scatter distributions. The high‐frequency scatter noise remains in the projection images even after a perfect scatter correction. In this paper, using a clinical CBCT system and a measurement‐based scatter correction, the authors show that a scatter correction alone does not provide satisfactory image quality and the loss of the contrast‐to‐noise ratio (CNR) of the scatter corrected image may overwrite the benefit of scatter removal. To circumvent the problem and truly gain from scatter correction, an effective scatter noise suppression method must be in place. They analyze the noise properties in the projections after scatter correction and propose to use a penalized weighted least‐squares (PWLS) algorithm to reduce the noise in the reconstructed images. Experimental results on an evaluation phantom (Catphan©600) show that the proposed algorithm further reduces the reconstruction error in a scatter corrected image from 10.6% to 1.7% and increases the CNR by a factor of 3.6. Significant image quality improvement is also shown in the results on an anthropomorphic phantom, in which the global noise level is reduced and the local streaking artifacts around bones are suppressed.