摘要
Fluvial sediments, including the NBS SRM-1645 Standard River Sediment, were fused in lithium tetraborate to form glass discs on which determinations of 16 elements (K, Ca, Ti, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, Rb, Sr, Y, Zr and Pb) were obtained simultaneously at a single set of x-ray conditions by energy-dispersive x-ray fluorescence spectrometry (e.d.x.r.f.). Relatively high sample-to-flux ratios of 1:3 were employed to obtained determinations of several minor as well as major elements on a single disc. Lower sample concentrations 1:6 and 1:10 were also used with the NBS SRM-1645. Inter-element corrections for absorption by iron were significant for detectable elements heavier than iron. In the NBS SRM-1645 samples, corrections were made for the fluorescence of chromium by the Fe K lines as well as the absorption of the Fe K lines by chromium. A background correction was also necessary for manganese in NBS SRM-1645. At the lowest sample-to-flux ratio (1:10) good agreement was obtained between the e.d.x.r.f. determinations and the NBS values for all quantifiable elements except zinc. At higher sample-to-flux ratios, agreement with the NBS values was generally poorer with increasing sample concentration. The relative standard deviation (RSD) of the eight quantifiable major and minor elements (K, Ca, Ti, Mn, Fe, Zn, Sr and Zr) determined under a single set of x-ray conditions ranged from 2 to 9% (RSD) with a mean RSD of 4.4% for a set of replica discs fused with Lake Pueblo/Arkansas River sediment samples.