热电效应
固相微萃取
材料科学
萃取(化学)
检出限
分析化学(期刊)
校准曲线
解吸
色谱法
固相萃取
大气温度范围
化学
质谱法
气相色谱-质谱法
吸附
热力学
物理
有机化学
作者
Yang Yang,Ángel Rodríguez-Lafuente,Janusz Pawliszyn
标识
DOI:10.1002/jssc.201400041
摘要
A temperature‐controlling device for in‐tube solid‐phase microextraction was developed based on thermoelectric cooling and heating. This device can control the temperature of the capillary column from 0 to 100°C by applying a voltage to a Peltier cooler or stainless steel tube. The extraction temperatures for angiotensin I, propranolol, and ranitidine were optimized. In all cases, setting the temperature to 10°C for extraction achieved the best extraction efficiency. Desorption showed minimum peak broadening at 70°C, contributing to better chromatographic performance. Propranolol was selected as a model compound to compare the performance of temperature‐controlled in‐tube solid‐phase microextraction at optimized conditions. Calibration curves exhibited good linearity ( R 2 > 0.999) over the studied range, and the limit of detection and limit of quantification were about three times lower than those obtained at standard conditions (30°C extraction and desorption).
科研通智能强力驱动
Strongly Powered by AbleSci AI