Pak Cheung Chan,Mary Rozmanc,Isolde Seiden‐Long,Jason Kwan
标识
DOI:10.1016/j.clinbiochem.2009.03.023
摘要
To evaluate the performance of the StatStrip™ glucose meter from Nova Biomedical for use in complex tertiary care facilities. Performance evaluation was conducted in 6 clinical locations involving nurse end-users. Imprecision (CV) < 5% was considered acceptable. Ten substances were examined for potential analytical interferences at two levels and 3 glucose concentrations. Inaccuracy was determined by 386 paired glucose differences between glucose meter and laboratory reference analyses. Acceptability was based on CSLI/ISO15197 criteria. Potential clinical impact was classified under a 5-level severity scheme. Imprecision varied from CV 2.4 to 4.9%. Significant interference was observed only for free hemoglobin at 10 g/L. 97% of paired glucose meter-laboratory differences met CSLI/ISO15197 criteria. All discrepant results had low severity scores. The StatStrip™ glucose meter demonstrated acceptable imprecision and inaccuracy, and was relatively free from common interferences. It should be a good candidate for general use in complex tertiary care facilities.