医学
血压
协议(科学)
手腕
人口
内科学
心脏病学
外科
病理
环境卫生
替代医学
作者
Zijian Xie,Bin Peng,Jia Hu,Xinda Wang,Xiaoqin Du,Chaoya Li,Jiahui Liang
出处
期刊:Blood Pressure Monitoring
[Ovid Technologies (Wolters Kluwer)]
日期:2024-09-13
标识
DOI:10.1097/mbp.0000000000000724
摘要
The objective of this study is to evaluate the accuracy of the wrist blood pressure (BP) monitor (TMB-2285-BT) in general population according to international standard of ISO 81060-2:2018+Amd.1:2020. The TMB-2285-BT BP monitor is an oscillometric device measuring BP from wrist. A single cuff for wrist sizes of 13.5–21.5 cm was used for the test device. The study was performed according to the ‘International Standard (ISO 81060-2:2018 + Amd.1:2020) protocol’ using the same-arm sequential BP measurement method. Subjects ( n ≥ 85) fulfilling the age, gender, BP, and cuff distribution criteria of the protocol were enrolled. For criterion 1, the mean difference of SBP and DBP between the test device and the reference device was −0.27 mmHg and −0.80 mmHg, respectively. For criterion 2, the SD of the averaged paired determinations of SBP and DBP for the test device and the reference device was 2.60 and 2.06 mmHg, respectively, both of which were less than 6.95 and 6.89 mmHg. The accuracy of the test device fulfilled the requirements of the criterion 1 and 2 of ISO 81060-2:2018 + Amd.1:2020. The TMB-2285-BT wrist BP monitor can be recommended for home BP measurement for general population.
科研通智能强力驱动
Strongly Powered by AbleSci AI