Xiao Zhang,Haozhe Zhong,Sainan Wang,Bin He,Liangqi Cao,Ming Li,Miaowen Jiang,Qin Li
标识
DOI:10.1002/jbio.202400104
摘要
A number of hardware-based and software-based strategies have been suggested to eliminate motion artifacts for improvement of 3D-optical coherence tomography (OCT) image quality. However, the hardware-based strategies have to employ additional hardware to record motion compensation information. Many software-based strategies have to need additional scanning for motion correction at the expense of longer acquisition time. To address this issue, we propose a motion artifacts correction and motion estimation method for OCT volumetric imaging of anterior segment, without requirements of additional hardware and redundant scanning. The motion correction effect with subpixel accuracy for in vivo 3D-OCT has been demonstrated in experiments. Moreover, the physiological information of imaging object, including respiratory curve and respiratory rate, has been experimentally extracted using the proposed method. The proposed method offers a powerful tool for scientific research and clinical diagnosis in ophthalmology and may be further extended for other biomedical volumetric imaging applications.