医学
扫描仪
置信区间
核医学
可靠性(半导体)
协议限制
平淡——奥特曼情节
计算机科学
人工智能
量子力学
物理
内科学
功率(物理)
作者
Shimpo Aoki,Hisato Nagano,Tomoharu Kiyosawa,Yoshifumi Tsuda,Ko Izumiyama
标识
DOI:10.1080/2000656x.2022.2032105
摘要
Acute hand edema often results in loss of fine hand motor activities, especially without appropriate care. There is still no reliable and easy to use method to measure hand edema. In this study, we tested a handheld three-dimensional (3D) scanner on plastic male and female hand models using a whole hand measuring method (WM) and a modified method (MM) which excluded fingers. We evaluated the intra-rater reliability and inter-rater reliability and compared the measured volumes to computed tomography (CT) findings. Statistical analysis showed that the 3D scan method was valid and reliable for both WM and MM methods. In WM, intra-rater and inter-rater reliability were 0.97 and 0.84, with 95% confidence interval (CI) of 0.87–1.00 and 0.61–0.94, respectively. In MM, intra-rater and inter-rater reliability were 0.96 and 0.83, with 95% CI of 0.84–1.00 and 0.61–0.94, respectively. In comparison to the CT, the differences between 3D scan and CT in the male model volumes were 30.35 ± 2.70 cm3 (mean ± standard deviation) for WM and 11.60 ± 2.07 cm3 for MM. In the female model, the differences were 18.92 ± 2.66 cm3 and 11.18 ± 2.35 cm3, respectively. In both models, MM was significantly more accurate than WM (p < 0.001). When used in a clinical case, the scanner recorded changes in actual volume through the course of treatment. This cost-effective handheld 3D camera can be a reliable tool for evaluating hand edema even in cases of acute injury.