医学
机制(生物学)
QRS波群
心脏病学
内科学
认识论
哲学
作者
Sudipta Mondal,Nayani Makkar,Narayanan Namboodiri
标识
DOI:10.1016/j.hrthm.2024.07.013
摘要
An electrocardiogram on the day following the implantation of a dual-chamber pacemaker revealed two closely coupled pacing spikes at 100ms intervals with the latter being on the trailing edge of the paced QRS complex. Differential diagnoses of such closely coupled pacing pulses include atrial lead displacement, lead reversal on the pulse generator, a ventricular premature complex in the post-atrial ventricular blanking period, and backup pacing pulse following a loss of capture. The critical analysis of the pacemaker timing intervals led to the mechanism and aetiology of the problem.