Hongchao Qi,Guangyin Zhang,Lin Xu,Yang Liu,Zhengzhi Wang,Yufu Xu,Ke Chen
标识
DOI:10.1021/acs.analchem.3c05084
摘要
A high-precision photoacoustic (PA) gas analyzer for fast dynamic measurement of ambient nitrogen dioxide (NO2) was developed. The PA analyzer used a differential PA cell combined with two mufflers to achieve rapid gas flow gas detection. A high-power laser diode (LD) with a center wavelength of 450 nm was used as the PA signal excitation source. To reduce the saturated absorption effect of NO2, ambient air was pumped into the analyzer at a flow rate of 900 sccm. Two mufflers were combined with the differential PA cell to reduce the noise caused by the airflow and pump. The parameters of the mufflers were optimized by using a finite element method. The experimental results showed that the gas flow noise was suppressed by 95%. The response time of the PAS analyzer was 34 s. The detection limits of the analyzer were 0.64 and 0.17 ppb when the integration times were 1 and 15 s, respectively. A 120 h continuous monitoring result was compared with the data from the National Environmental Monitoring Station to demonstrate the high reliability of the analyzer.