Min Joo Choi,Sitaramanjaneya Reddy Guntur,Joo Myoung Lee,Dong Guk Paeng,Kang I.L. Lee,Andrew Coleman
标识
DOI:10.1016/j.ultrasmedbio.2011.06.015
摘要
The present work considers the ultrasonic properties of porcine liver tissue in vitro measured during heating concomitant with thermal coagulation followed by natural cooling, so as to provide information about changes in the ultrasonic properties of the tissue after thermal coagulation. The excised liver samples were heated in a degassed water bath up to 75°C and naturally cooled down to 30°C. The tissue was observed to begin thermally coagulating at temperatures lower than 75°C. The ultrasonic parameters considered include the speed of sound, the attenuation coefficient, the backscatter coefficient and the nonlinear parameter of B/A. They were more sensitive to temperature when heating than during natural cooling. All of the parameters were shown to rise significantly on completion of the heating–cooling cycle. At 35°C after thermal coagulation, the B/A value was increased by 96%, the attenuation and backscatter coefficients were increased by 50%∼68% and 33%∼37%, respectively, in the typical frequency ranges of 3 MHz∼5 MHz used for ultrasonic imaging and the speed of sound was increased by 1.4%. The results of this study added to the evidence that tissue characterization, in particular, based on the B/A could be valuable for ultrasonically imaging the thermal lesions following high-intensity focused ultrasound (HIFU) surgery.