Robert M. Sturm,Barry R Jones,Daniel Mulvana,Steve Lowes
出处
期刊:Bioanalysis
[Newlands Press Ltd]
日期:2016-08-01
卷期号:8 (16): 1709-1721
被引量:24
标识
DOI:10.4155/bio-2016-0079
摘要
High-resolution MS (HRMS) has seen an uptake in use for discovery qual/quan workflows, however, its utilization in late discovery/development has been slow. Past reports comparing HRMS to triple quadrupole (QQQ) instrumentation to date have indicated that HRMS instruments are capable of producing data acceptable for regulated bioanalysis, however lack the sensitivity required for sub ng/ml LLOQ assays. Recent advances in HRMS instrumentation have closed the sensitivity gap with QQQ and have even provided improved selectivity and sensitivity over QQQ SRM assays. Herein, the authors will describe how, when, and why HRMS (specifically Q-Exactive series mass spectrometers) should be considered for implementation in regulated quantitative bioanalysis assays.