期刊:IEEE Sensors Journal [Institute of Electrical and Electronics Engineers] 日期:2024-01-01卷期号:: 1-1
标识
DOI:10.1109/jsen.2024.3398617
摘要
This study introduces a successfully manufactured U-shaped optical fiber sensor integrated with a microfluidics system to detect varying glucose concentrations. The outlines the fabrication process of the microfluidics system, accomplished using a CO2 laser on polymethyl methacrylate (PMMA) material. Through three cycles of experiments involving different glucose concentrations, the results indicate a consistent trend in both resonance wavelength and transmission loss, showing an increase as glucose concentration rises from 5% to 50%. The system demonstrates a high wavelength sensitivity coefficient of 0.114 nm/% with R 2 = 0.95 and a notable transmission loss sensitivity of 0.09 dB/% and R 2 = 0.94. This advanced system proves to be effective in achieving rapid, sensitive, and cost-efficient glucose monitoring. Consequently, the microfluidics system is deemed suitable for its intended purpose.