Lipids serve an essential role in multiple cellular functions including signaling, metabolism, energy storage, and membrane constitution. Lipidomics, the study of lipids using analytical chemistry, allows for the study of disease states and cellular metabolism. Shotgun lipidomics is a technique that involves direct-infusion electrospray ionization (ESI) and analysis with a triple quadrupole mass spectrometer. Triple quadrupole mass spectrometry is ideally suited for lipidomics analysis because it allows for class-specific identification of lipids. Individual lipid class can be identified by the adjustment of three parameters-collision energy, ion mode, and scan type. This chapter describes the use of a triple quadrupole mass spectrometer, the TSQ Quantum Access MAX, to perform lipidomics analysis with high sensitivity, accuracy, and precision.