Xiuyuan Li,Pingjun Chang,Zhangliang Li,Shuyi Qian,Zehui Zhu,Qianwei Wang,Ying‐Zheng Zhao
标识
DOI:10.1080/17434440.2020.1848541
摘要
Purpose: To evaluate the agreement between anew UBM and an SS-OCT.Methods: The scans of the right eye of each volunteer were obtained using the two devices. Data were fitted and recorded including: central corneal thickness (CCT), aqueous depth (AQD) (the distance from endothelium to lens), angle-to-angle distance (ATA), lens thickness (LT), diameter of the lens in the horizontal direction (LDiaangle: distance between the sharp angles on both sides of the lens, LDiaarc: distance between the vertex of the circular arcs on both sides of the lens), anterior and posterior corneal radius (Rf and Rb).Results: 25 eyes were included in this study. It could be seen that the differences in CCT, LDiaangle, Rf measured by the two instruments were not statistically significant. Bland-Altman analysis plots of CCT, LDiaangle and Rf showed mean differences of 0.2 µm, 0.01mm and 0.0mm for the 2 devices, respectively.Conclusion: The values of CCT, LDiaangle and Rf obtained via two instruments were not clinically interchangeable and the AQD, ATA, LT, and Rb have poor agreement affected by accommodation. We can estimate the real lens diameter by subtracting 0.61 ± 0.43mm when the lens diameter can only be simulated with SS-OCT.