材料科学
电解质
分析器
安培法
生物医学工程
电极
氧传感器
气体分析仪
氧气
工艺工程
分析化学(期刊)
化学
电化学
色谱法
工程类
环境化学
物理化学
有机化学
作者
Yu-Zhi Lam,John Atkinson
标识
DOI:10.1016/j.medengphy.2006.03.010
摘要
Transcutaneous blood gas monitoring is a non-invasive measurement technique for obtaining fast and relatively accurate responses to determine the respiratory conditions of patients. In this investigation, a screen-printed, disposable, transcutaneous oxygen sensor based on the working principle of amperometry, incorporates an integral heating element to enhance transcutaneous diffusion of blood gases typically at 44 degrees C. A Clark cell configuration is employed with gold working and counter electrodes and a silver/silver chloride reference electrode. Two different types of electrolytes namely potassium nitrate gel and polyelectrolyte (Nafion) have been studied under laboratory test conditions. A fully computer-controlled gas testing rig has been constructed to automate the varying of oxygen levels. Linear relationships have been established with an averaged sensitivity level of 0.029 microA/mmHg. In addition, a brief pilot clinical trial involving a fully grown human subject has been carried out alongside a commercial transcutaneous blood gas analyser. The investigations have shown that although the measured signals are weaker than those obtained from the laboratory test, the thick film sensor displays a repeatable and linear relationship when correlating with the commercial system. This study has greatly contributed towards the understanding for the suitability of the materials in achieving a viable, low-cost biomedical sensor.
科研通智能强力驱动
Strongly Powered by AbleSci AI