Accuracy of arthroscopic fluid pump systems in shoulder surgery: a comparison of 3 different pump systems

医学 关节镜检查 外科 外渗 压力传感器 麻醉 热力学 物理 免疫学
作者
Mohy E. Taha,K. Schneider,Margaret M. Smith,Grégory Cunningham,Allan Young,Benjamin Cass
出处
期刊:Journal of Shoulder and Elbow Surgery [Elsevier BV]
卷期号:29 (12): 2626-2631 被引量:6
标识
DOI:10.1016/j.jse.2020.03.039
摘要

Background Extra-articular fluid extravasation is a known complication during shoulder arthroscopy. The risk and amount of extravasation to a large degree is dependent on the fluid pressure delivered to the surgical site. Accurate measurement, knowledge, and control of the pressure delivered is thus important to surgeons, anesthetists, and the patient. The purpose of this study was to compare the pressure measurement accuracy of 3 arthroscopic fluid pumps, with 2 of them having 2 different settings. Methods Twenty-five patients (n = 5 per group) undergoing shoulder arthroscopy were selected. Three different arthroscopic fluid pumps (ConMed 24K, Stryker Crossflow, Arthrex Dual Wave) were tested in 5 different operational settings (Stryker, standard and dynamic mode; ConMed, with and without TIPS; Arthrex Dual Wave). In each operation, the set pump pressures and the subsequently delivered intra-articular surgical site fluid pressures were measured by a spinal needle connected to an anesthetic standard pressure transducer attached to the anesthetic machine. Independent measures of the surgical site pressures were obtained before multiple portals were created or extravasation had occurred. Measurements were taken at the beginning of surgery. Results Measurements of the mean intra-articular pressure were found to not be significantly different from the set pressure for the ConMed 24K with TIPS (0.98 ± 0.02-fold) and Stryker Crossflow in standard mode (0.98 ± 0.02-fold). However, actual pressure was significantly greater than the set pressure for the ConMed 24K without TIPS (by 1.30 ± 0.13-fold), Stryker Crossflow in dynamic mode (by 1.82 ± 0.08-fold), and Arthrex Dual Wave (by 2.19 ± 0.06-fold). Conclusion Independently measured intra-articular pressure can be more than double the set pressure for some arthroscopic pumps. Measuring intra-articular pressure can thus aid in adjusting the set pressure. This could minimize the risk of intraoperative complications. Extra-articular fluid extravasation is a known complication during shoulder arthroscopy. The risk and amount of extravasation to a large degree is dependent on the fluid pressure delivered to the surgical site. Accurate measurement, knowledge, and control of the pressure delivered is thus important to surgeons, anesthetists, and the patient. The purpose of this study was to compare the pressure measurement accuracy of 3 arthroscopic fluid pumps, with 2 of them having 2 different settings. Twenty-five patients (n = 5 per group) undergoing shoulder arthroscopy were selected. Three different arthroscopic fluid pumps (ConMed 24K, Stryker Crossflow, Arthrex Dual Wave) were tested in 5 different operational settings (Stryker, standard and dynamic mode; ConMed, with and without TIPS; Arthrex Dual Wave). In each operation, the set pump pressures and the subsequently delivered intra-articular surgical site fluid pressures were measured by a spinal needle connected to an anesthetic standard pressure transducer attached to the anesthetic machine. Independent measures of the surgical site pressures were obtained before multiple portals were created or extravasation had occurred. Measurements were taken at the beginning of surgery. Measurements of the mean intra-articular pressure were found to not be significantly different from the set pressure for the ConMed 24K with TIPS (0.98 ± 0.02-fold) and Stryker Crossflow in standard mode (0.98 ± 0.02-fold). However, actual pressure was significantly greater than the set pressure for the ConMed 24K without TIPS (by 1.30 ± 0.13-fold), Stryker Crossflow in dynamic mode (by 1.82 ± 0.08-fold), and Arthrex Dual Wave (by 2.19 ± 0.06-fold). Independently measured intra-articular pressure can be more than double the set pressure for some arthroscopic pumps. Measuring intra-articular pressure can thus aid in adjusting the set pressure. This could minimize the risk of intraoperative complications.

科研通智能强力驱动
Strongly Powered by AbleSci AI
科研通是完全免费的文献互助平台,具备全网最快的应助速度,最高的求助完成率。 对每一个文献求助,科研通都将尽心尽力,给求助人一个满意的交代。
实时播报
FFF发布了新的文献求助20
1秒前
壮观冷卉发布了新的文献求助10
2秒前
爆米花应助Jessie采纳,获得10
3秒前
深情安青应助会飞的猪采纳,获得10
3秒前
4秒前
山茶完成签到,获得积分10
4秒前
AKira发布了新的文献求助10
6秒前
7秒前
温暖香菱发布了新的文献求助10
7秒前
周周完成签到 ,获得积分10
7秒前
小确幸完成签到,获得积分10
7秒前
顾矜应助1asfdwe采纳,获得10
7秒前
嗯呐完成签到,获得积分10
7秒前
蓝天发布了新的文献求助10
8秒前
Cetus发布了新的文献求助10
9秒前
怕孤独的鹭洋完成签到,获得积分10
9秒前
10秒前
今后应助YXL采纳,获得30
11秒前
wyt123发布了新的文献求助10
12秒前
天天快乐应助哈哈蔡酱采纳,获得10
13秒前
13秒前
laber完成签到,获得积分0
14秒前
14秒前
123发布了新的文献求助10
15秒前
撒上大声说完成签到,获得积分10
17秒前
17秒前
GDcomeback完成签到,获得积分10
17秒前
19秒前
FashionBoy应助Rose采纳,获得10
20秒前
王多肉完成签到,获得积分10
20秒前
充电宝应助kuku_99采纳,获得10
20秒前
Nina完成签到,获得积分10
22秒前
helly完成签到,获得积分10
22秒前
吴晨曦发布了新的文献求助10
23秒前
xinyi发布了新的文献求助10
24秒前
眼睛大的可乐完成签到,获得积分10
24秒前
梦月无声发布了新的文献求助10
24秒前
zz完成签到,获得积分10
25秒前
25秒前
xiaolengyan发布了新的文献求助10
25秒前
高分求助中
(应助此贴封号)【重要!!请各用户(尤其是新用户)详细阅读】【科研通的精品贴汇总】 10000
The Wiley Blackwell Companion to Diachronic and Historical Linguistics 3000
The impact of workplace variables on juvenile probation officers’ job satisfaction 1000
When the badge of honor holds no meaning anymore 1000
HANDBOOK OF CHEMISTRY AND PHYSICS 106th edition 1000
ASPEN Adult Nutrition Support Core Curriculum, Fourth Edition 1000
AnnualResearch andConsultation Report of Panorama survey and Investment strategy onChinaIndustry 1000
热门求助领域 (近24小时)
化学 材料科学 医学 生物 纳米技术 工程类 有机化学 化学工程 生物化学 计算机科学 物理 内科学 复合材料 催化作用 物理化学 光电子学 电极 细胞生物学 基因 无机化学
热门帖子
关注 科研通微信公众号,转发送积分 6280622
求助须知:如何正确求助?哪些是违规求助? 8099772
关于积分的说明 16934222
捐赠科研通 5348168
什么是DOI,文献DOI怎么找? 2842909
邀请新用户注册赠送积分活动 1820285
关于科研通互助平台的介绍 1677197