光学
中央凹
自适应光学
扫描激光检眼镜
人眼
图像分辨率
物理
计算机科学
眼科
医学
视网膜
作者
Benjamin Moon,Martina Poletti,Austin Roorda,Pavan Tiruveedhula,Soh Hang Liu,Glory Linebach,Michele Rucci,Jannick P. Rolland
摘要
In prior art, advances in adaptive optics scanning laser ophthalmoscope (AOSLO) technology have enabled cones in the human fovea to be resolved in healthy eyes with normal vision and low to moderate refractive errors, providing new insight into human foveal anatomy, visual perception, and retinal degenerative diseases. These high-resolution ophthalmoscopes require careful alignment of each optical subsystem to ensure diffraction-limited imaging performance, which is necessary for resolving the smallest foveal cones. This paper presents a systematic and rigorous methodology for building, aligning, calibrating, and testing an AOSLO designed for imaging the cone mosaic of the central fovea in humans with cellular resolution. This methodology uses a two-stage alignment procedure and thorough system testing to achieve diffraction-limited performance. Results from retinal imaging of healthy human subjects under 30 years of age with refractive errors of less than 3.5 diopters using either 680 nm or 840 nm light show that the system can resolve cones at the very center of the fovea, the region where the cones are smallest and most densely packed.
科研通智能强力驱动
Strongly Powered by AbleSci AI