化学
色谱法
液相色谱-质谱法
串联质谱法
串联
人血浆
质谱法
复合材料
材料科学
作者
Hye Young Ji,Dong Won Jeong,Young Hoon Kim,Hui‐Hyun Kim,Y. Yoon,Kang Choon Lee,Hye Suk Lee
摘要
Abstract A rapid, sensitive and selective method for the determination of gabapentin in human plasma was developed using hydrophilic interaction liquid chromatography/tandem mass spectrometry (HILIC/MS/MS). The devised method involved protein precipitation with acetonitrile followed by separation on an Atlantis HILIC silica column using an acetonitrile/ammonium formate mobile phase (100 mM, pH 3.0) (85:15, v/v). Analytes were detected using an electrospray ionization mass spectrometer in the multiple‐reaction monitoring mode. The standard curve was linear ( r = 1.000) over the concentration range of 50.0–10000 ng/mL. The lower limit of quantification for gabapentin was 50.0 ng/mL (ca. 20 pg gabapentin) using a 10‐µL plasma sample. The coefficients of variation and relative errors for intra‐ and inter‐assay at four QC levels (i.e., 50.0, 125, 750, and 7500 ng/mL) were 4.7 to 9.4% and −4.1 to 1.6%, respectively. Absolute and relative matrix effects for gabapentin and metformin were practically absent. Gabapentin and metformin recoveries were 98.5% and 99.0%, respectively. This method was successfully applied to a bioequivalence study of gabapentin in humans. Copyright © 2006 John Wiley & Sons, Ltd.