Yongde Wu,Xian-Zhong Mei,Wei-Fei Wu,Hongxi Zhang,Jie Liang,Xianhua Cai
标识
DOI:10.1186/s12891-024-07646-0
摘要
Abstract Background and objective Complex acetabular fractures involving quadrilateral areas are more challenging to treat during surgery. To date, there has been no ideal internal fixation for these acetabular fractures. The purpose of this study was to evaluate the biomechanical stability of complex acetabular fractures using a dynamic anterior titanium-plate screw system of the quadrilateral area (DAPSQ) by simulating the standing and sitting positions of pelvic specimens. Materials and methods Eight formal in-preserved cadaveric pelvises aged 30–50 years were selected as the research objects. First, one hip of the normal pelvises was randomly used as the control model (group B) for measurement, and then one hip of the pelvises was randomly selected to make the fracture model in the 8 intact pelvises as the experimental model (group A) for measurement. In group A, acetabular both-column fractures in the quadrilateral area were established, and the fractures were fixed by DAPSQ. The biomechanical testing machine was used to load (simulated physiological load) from 400 N to 700 N at a 1 mm/min speed for 30 s in the vertical direction when the specimens were measured at random in simulated standing or sitting positions in groups. The horizontal displacement and longitudinal displacement of the acetabular fractures in the quadrilateral area were measured in both the standing and sitting simulations. Results As the load increased, no dislocation or internal fixation breakage occurred during the measurements. In the standing position, the horizontal displacement of the quadrilateral area fractures in group A and group B appeared to be less than 1 mm with loads ranging from 400 N to 700 N, and there was no significant difference between group A and group B ( p > 0.05). The longitudinal displacement appeared to be greater than 1 mm with a load of 700 mm in group A (700 N, 2 cases), and the difference was significant between group A and group B ( p < 0.05). In the sitting position, the horizontal and longitudinal displacements of the quadrilateral areas were within 0.5 mm in group A and group B, and there was no significant difference between group A and group B ( p > 0.05). Conclusion For complex acetabular fractures in the quadrilateral area, DAPSQ fixation may provide early sitting stability, but it is inappropriate for patients to stand too early.
科研通智能强力驱动
Strongly Powered by AbleSci AI