Boris Jegorović,Neda Milinković,Marija Sarić Matutinović,Sandra Šipetić,Svetlana Ignjatović
标识
DOI:10.1080/00365513.2022.2150983
摘要
This study aimed to examine the agreement of the serum amyloid A (SAA) values determined using the ELISA test and the nephelometric automated method. This study included 80 serum samples obtained from patients with COVID-19. Samples were determined using ELISA and the nephelometric method. Wilcoxon signed ranks test showed a statistically significant difference in the calculated median values (Z = -2.432, p = 0.015). The correlation between methods was statistically significant (r = 0.603, p < 0.0001). Bland Altman analysis showed a bias of 56.6 mg/L and a relative bias of 7.4% between the methods. The results of this study indicate that further studies are needed that will examine the compliance between the ELISA and the nephelometric method for determining SAA, and the results must be carefully interpreted based on the method used.