人工晶状体度数计算
数学
人工晶状体
功率(物理)
应用数学
光学
角膜
物理
角膜曲率计
热力学
作者
John Retzlaff,Donald R. Sanders,Manus C. Kraff
出处
期刊:Journal of Cataract and Refractive Surgery
[Ovid Technologies (Wolters Kluwer)]
日期:1990-05-01
卷期号:16 (3): 333-340
被引量:934
标识
DOI:10.1016/s0886-3350(13)80705-5
摘要
A new implant power calculation formula (SRK/T) was developed using the nonlinear terms of the theoretical formulas as its foundation but empirical regression methodology for optimization. Postoperative anterior chamber depth prediction, retinal thickness axial length correction, and corneal refractive index were systematically and interactively optimized using an iterative process on five data sets consisting of 1,677 posterior chamber lens cases. The new SRK/T formula performed slightly better than the Holladay, SRK II, Binkhorst, and Hoffer formulas, which was the expected result as any formula performs superiorly with the data from which it was derived. Comparative accuracy of this formula upon independent data sets is addressed in a follow-up report. The formula derived provides a primarily theoretical approach under the SRK umbrella of formulas and has the added advantage of being calculable using either SRK A-constants that have been empirically derived over the last nine years or using anterior chamber depth estimates.
科研通智能强力驱动
Strongly Powered by AbleSci AI