Rémi Philippot,Jean François Meucci,Bertrand Boyer,F. Farizon
出处
期刊:PubMed
日期:2013-09-01
卷期号:23: 208-12
被引量:25
摘要
We report the results of a 12-year follow-up retrospective series of 100 total hip arthroplasties using cementless, press-fit, dual-mobility acetabular cups. The aim of our study was to evaluate the clinical and radiographic results of this acetabular cup at last follow-up. This continuous and homogeneous series included 100 primary total hip arthroplasties performed during the year 2000. The THA combined a Corail® stem (Corail®, Depuy, Warsaw, IN) with a stainless steel Novae Sunfit® (Serf, Decines, France) acetabular cup. Fifteen patients died and 2 were lost to follow-up. Two cases of early dislocation were observed, and 3 cases of aseptic loosening of the acetabular component were reported. The mean stem subsidence was 0.71 mm, the mean craniopodal acetabular migration was 1.37 mm, and the mean medio-lateral acetabular migration was 1.52 mm. The 12-year survivorship is comparable to the data from the literature. The low dislocation rate at 12 years confirms the long-term, high stability of dual mobility, which should be recommended in primary THA for patients at risk for postoperative instability. The absence of true intraprosthetic dislocation events at 12-year follow-up provides evidence of the good quality of the latest generation of polyethylene liners and the necessity of combining thin-mirror, polished femoral necks with dual-mobility cups.