光学
干涉测量
折射率
棱镜
材料科学
波长
傅里叶变换光谱学
天文干涉仪
激光器
傅里叶变换红外光谱
物理
作者
Yukio Hori,Akiko Hirai,Kaoru Minoshima
出处
期刊:Applied Optics
[The Optical Society]
日期:2014-04-24
卷期号:53 (13): 2793-2793
被引量:4
摘要
A prism-pair interferometer for a spectrum lamp was developed for precise measurement of the refractive index of a prism of optical glass. Previously we reported the prism-pair interferometer with a He–Ne laser light source, resulting in a measurement uncertainty of 1.1×10−6. However, most of the refractive-index values managed by optical glass manufacturers are conventionally measured with spectrum lamps. We have optimized the prism-pair interferometer for spectrum lamps and implemented a signal-processing technique from Fourier-transform spectroscopy. When the refractive index is measured, the wavelength of the spectrum lamp is simultaneously calibrated by part of the interferometer, so that the resulting refractive index is traceable to a national standard of length. The combined standard uncertainty for a refractive index measured with the e-line (546 nm) of a Hg lamp is 6.9×10−6.