Ramin Khoramnia,Isabella D. Baur,Grzegorz Łabuz,Maximilian K. Köppe,Mustafa K. Hallak,Gerd U. Auffarth
出处
期刊:Journal of Cataract and Refractive Surgery
[Ovid Technologies (Wolters Kluwer)]
日期:2023-10-01
卷期号:49 (10): 1011-1017
被引量:5
标识
DOI:10.1097/j.jcrs.0000000000001265
摘要
To clinically evaluate a diffractive continuous-range-of-vision intraocular lens (IOL) that combines bifocal and extended depth-of-focus technologies in refractive lens exchange (RLE) patients.University eye clinic.Prospective interventional clinical study.Bilateral implantation of the TECNIS Synergy IOL was performed during a RLE procedure in 28 patients. At 3 months postoperatively, uncorrected (UDVA) and corrected (CDVA) distance visual acuities, uncorrected (UIVA) and distance-corrected (DCIVA) intermediate visual acuities at 80 cm, and uncorrected (UNVA) and distance-corrected (DCNVA) near visual acuities at 40 cm were assessed. Defocus curve testing and contrast sensitivity testing were also performed.Binocular UDVA and CDVA of 56 eyes (28 patients) were -0.06 ± 0.07 logMAR and -0.13 ± 0.05 logMAR, respectively. UIVA and DCIVA were -0.06 ± 0.05 logMAR and -0.08 ± 0.06 logMAR, and UNVA and DCNVA were -0.01 ± 0.07 logMAR and -0.04 ± 0.07 logMAR. The defocus curve revealed a visual acuity of 0.10 logMAR or better from +0.50 to -3.00 diopters (D). Photopic contrast sensitivity at spatial frequencies 3.0, 6.0, 12.0, and 18.0 was 1.68 ± 0.20, 1.77 ± 0.19, 1.45 ± 0.18, and 1.02 ± 0.19 log units, respectively. Mesopic contrast sensitivity at the same frequencies was 1.56 ± 0.20, 1.44 ± 0.25, 0.92 ± 0.38, and 0.44 ± 0.37, respectively.The Synergy IOL provided very good distance, intermediate, and near visual outcomes. A visual acuity of 0.10 logMAR or better was achieved from +0.50 to -3.0 D. Contrast sensitivity was within the normal range.
科研通智能强力驱动
Strongly Powered by AbleSci AI