医学
铅(地质)
背景(考古学)
重症监护医学
植入式心律转复除颤器
心脏病学
生物
地貌学
地质学
古生物学
作者
Charles D. Swerdlow,Gautham Kalahasty,Kenneth A. Ellenbogen
标识
DOI:10.1016/j.jacc.2015.12.067
摘要
The implantable-cardioverter defibrillator (ICD) lead is the most vulnerable component of the ICD system. Despite advanced engineering design, sophisticated manufacturing techniques, and extensive bench, pre-clinical, and clinical testing, lead failure (LF) remains the Achilles’ heel of the ICD system. ICD LF has a broad range of adverse outcomes, ranging from intermittent inappropriate pacing to proarrhythmia leading to patient mortality. ICD LF is often considered in the context of design or construction defects, but is more appropriately considered in the context of the finite service life of a mechanical component placed in chemically stressful environment and subjected to continuous mechanical stresses. This clinical review summarizes LF mechanisms, assessment, and differential diagnosis of LF, including lead diagnostics, recent prominent lead recalls, and management of LF and functioning, but recalled leads. Despite recent advances in lead technology, physicians will likely continue to need to understand how to manage patients with transvenous ICD leads.
科研通智能强力驱动
Strongly Powered by AbleSci AI