肌酐
分析物
注意事项
检出限
RGB颜色模型
比色法
尿
线性范围
生物医学工程
色调
计算机科学
人工智能
色谱法
化学
数学
医学
生物化学
病理
作者
Izabela Lewińska,Mikołaj Speichert,Mateusz Granica,Łukasz Tymecki
标识
DOI:10.1016/j.snb.2021.129915
摘要
Creatinine is a clinically significant analyte used to diagnose kidney condition. However, the literature still lacks in creatinine sensors fulfilling point-of-care testing requirements. In this paper, we have developed colorimetric paper-based creatinine sensors adhering to point-of-care testing principles. The signal readout is accomplished with a smartphone modified with 3D-printed elements and processed with a self-written application compromising computer vision algorithm for automatic detection of the colored zone. Two colorimetric methods – routinely used Jaffé method and an alternative one with 3,5-dinitrobenzoic acid were both tested and compared. Hue channel intensity from HSV color space and green channel from RGB color space was used as the analytical signal in Jaffé and 3,5-dinitrobenzoate method, respectively. For both kinds of sensors the linear range of the response covered the range significant for urinary analysis, with precision, expressed as RSD, below 5%. Limit of quantification for Jaffé method was 1.05 mmol·L−1 whereas it was 0.82 mmol·L−1 for 3,5-dinitrobenzoate method. The utility of the developed sensors to selectively quantify creatinine in undiluted urine was proved using artificial urine samples and the obtained recoveries were in the range from 70 to 129 %.