圆窗
人工耳蜗植入术
听力学
人工耳蜗植入
窗口(计算)
插入损耗
医学
材料科学
耳蜗
计算机科学
光电子学
操作系统
作者
Oliver F. Adunka,Marc H. Unkelbach,Martin G. Mack,Markus Hambek,Wolfgang Gstöettner,Jan Kiefer
标识
DOI:10.1080/00016480410018179
摘要
Objective To evaluate cochlear implant trauma to intracochlear structures when inserting the electrode via the round window membrane. Material and methods Eight fresh human temporal bones were evaluated histologically after insertion using two types of cochlear implant array. Bones underwent a special fixation and embedding procedure that allowed sectioning of undecalcified bone with the electrode in situ. Insertions depths were evaluated radiologically and histologically. Results All arrays were found in the scala tympani of the cochlea. Basal trauma could be avoided in all but one specimen. The mean depth of insertion was 382.5°. Apically, only one implanted bone showed cochlear trauma exceeding lifting of the basilar membrane. Conclusion Insertions through the round window membrane were shown to be atraumatic, even in basal cochlear regions. This route of insertion might be very effective for combined electric and acoustic stimulation of the auditory system.
科研通智能强力驱动
Strongly Powered by AbleSci AI