医学
心室流出道
经皮
肺动脉瓣
心脏病学
主动脉瓣
内科学
心脏瓣膜
支架
外科
作者
Patcharapong Suntharos,Lourdes Prieto
标识
DOI:10.1177/2150135116655125
摘要
Percutaneous pulmonary valve implantation is now an acceptable treatment option. The Edwards SAPIEN valve, intended for transcatheter aortic valve implantation, has been used in patients whose landing zone diameter is >22 to 24 mm which is not suitable for Melody valve. We report a patient in whom we successfully placed the latest generation, Edwards SAPIEN 3 (ES3) in the pulmonary position. However, because of the short covered portion of this valve, there was a significant paravalvular leak through the stent struts requiring implantation of a second ES3 valve.