分光计
校准
发射光谱
钬
镧系元素
直线(几何图形)
能量(信号处理)
原子物理学
铽
材料科学
物理
计算物理学
光学
谱线
激光器
发光
离子
天文
量子力学
数学
几何学
作者
Joseph W. Fowler,G. C. O’Neil,Bradley K. Alpert,D. A. Bennett,E. V. Denison,W. B. Doriese,Gene C. Hilton,Lawrence T. Hudson,Y-I Joe,Kelsey M. Morgan,D. R. Schmidt,Daniel S. Swetz,Csilla I. Szabo,Joel N. Ullom
出处
期刊:Metrologia
[IOP Publishing]
日期:2020-12-11
卷期号:58 (1): 015016-015016
被引量:24
标识
DOI:10.1088/1681-7575/abd28a
摘要
We use an array of transition-edge sensors, cryogenic microcalorimeters with 4 eV energy resolution, to measure L x-ray emission-line profiles of four elements of the lanthanide series: praseodymium, neodymium, terbium, and holmium. The spectrometer also surveys numerous x-ray standards in order to establish an absolute-energy calibration traceable to the international system of units for the energy range 4 keV to 10 keV. The new results include emission line profiles for 97 lines, each expressed as a sum of one or more Voigt functions; improved absolute energy uncertainty on 71 of these lines relative to existing reference data; a median uncertainty on the peak energy of 0.24 eV, four to ten times better than the median of prior work; and six lines that lack any measured values in existing reference tables. The 97 lines comprise nearly all of the most intense L lines from these elements under broad-band x-ray excitation. The work improves on previous measurements made with a similar cryogenic spectrometer by the use of sensors with better linearity in the absorbed energy and a gold x-ray absorbing layer that has a Gaussian energy-response function. It also employs a novel sample holder that enables rapid switching between science targets and calibration targets with excellent gain balancing. Most of the results for peak energy values shown here should be considered as replacements for the currently tabulated standard reference values, while the line shapes given here represent a significant expansion of the scope of available reference data.
科研通智能强力驱动
Strongly Powered by AbleSci AI