期刊:Journal of Drugs in Dermatology [SanovaWorks] 日期:2020-11-01卷期号:19 (12): 1199-1203被引量:20
标识
DOI:10.36849/jdd.2020.5228
摘要
Background Poly-L-lactic acid (PLLA) is a biodegradable, synthetic polymer that stimulates collagen production and can improve skin quality, volume, and thickness. The current reconstitution procedure for Sculptra, a PLLA-containing injectable device involves 2 hours standing time before use. Objective To evaluate and validate an immediate-use procedure for reconstituting a PLLA-containing injectable device. Methods and materials Three batches of the product were shaken for 1 minute immediately after reconstitution with 8 mL of sterile water. Different physicochemical tests including viscosity, concentration of excipients (sodium carboxymethylcellulose and mannitol), pH, and particle size distribution were performed for standing times 0, 2, 24, and 72 hours after immediate shaking, and compared with the standard 2 hours standing time before shaking. The recovery and stability of optional addition of 1 mL of 2% lidocaine hydrochloride was also assessed. Results All physiochemical parameters evaluated were equivalent, regardless of reconstitution procedure, showing that shaking vigorously for 1 minute dissolves the excipients of the product properly without a required standing time and with no impact to the PLLA particles. There were no differences in lidocaine hydrochloride content of suspensions after 0 and 72 hours. Conclusion The PLLA-containing product can be used immediately after reconstitution including vigorous shaking, as shown from physicochemical analyses. Optional addition of lidocaine hydrochloride is feasible. J Drugs Dermatol. 2020;19(12): doi:10.36849/JDD.2020.5228.