介电谱
蛋白质吸附
循环伏安法
生物污染
吸附
化学
牛血清白蛋白
色谱法
生物传感器
电化学
生物化学
电极
膜
物理化学
有机化学
作者
Matthew F. Smiechowski,Vadim F. Lvovich,Shuvo Roy,Aaron J. Fleischman,William H. Fissell,Alan T. Riga
标识
DOI:10.1016/j.bios.2006.02.008
摘要
On-line detection of serum proteins is of clinical relevance, in detecting leaks and biofouling in hemofiltration equipment, biofilm growth on prosthetic devices, or hemolysis within a prosthetic or therapeutic device. Electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV) were employed to detect and analyze micromolar concentrations of four globular proteins of clinical importance. CV testing showed that identification and quantification of each of these proteins was possible through analysis of current changes at specific potentials. Preliminary CV studies into the contamination of Bovine Serum Albumin with a microgram amount of one of the other three proteins illustrated that direct detection of the contaminant protein was possible. The analysis of the EIS data demonstrated that with increase in relative concentration of proteins, the amount of electroactive proteins adsorption at the interface increases, leading to increase in surface charge density and capacitance, especially for lower molecular weight proteins. The impedance data was used to determine the values of Gibbs adsorption energy, adsorption coefficients for the four proteins, and develop an equivalent circuit model for the protein-containing solutions.
科研通智能强力驱动
Strongly Powered by AbleSci AI