Tengfei Bao,Yin Liu,Caihui Liu,Yunlei Pang,Guisen Zhang,Yajun Yang,Bozhou Zhang
标识
DOI:10.1016/j.pdpdt.2022.103207
摘要
To compare the inter-device agreement of anterior eye segment measurements between LenStar LS 900 optical biometer and OPD Scan III wavefront aberrometer. This is a retrospective study involving 59 patients (78 eyes) with cataract. Their angle Alpha, angle Kappa, pupil size and white-to-white (WTW) distance were measured by LenStar LS 900 optical biometer and OPD Scan III wavefront aberrometer, respectively, and pairwise agreement comparisons were performed between them. The most agreement of various parameters was occurred, with intraclass correlation coefficient (ICC) of WTW = 0.930; angle Alpha = 0.853; angle Kappa = 0.898; and pupil size = 0,976 in bright environment. Furthermore, in dark environment, the ICC of WTW, angle Alpha, angle Kappa, and pupil size were 0.927, 0.791, 0.915, and 0.990, respectively. Bland-Altman plot showed similar excellent agreement in the outcomes of the two devices for these measurements testing. There was an excellent agreement between the LenStar LS 900 optical biometer and OPD Scan III wavefront aberrometer for WTW, angle Alpha, angle Kappa, and pupil size measurements. In clinical practice, these measurements obtained by LenStar LS 900 optical biometer and OPD Scan III wavefront aberrometer can be used interchangeably.