材料科学
制作
超声波传感器
传感器
部分带宽
电容感应
电容式微机械超声换能器
共形映射
超声波
曲率半径
表面微加工
曲率
带宽(计算)
光电子学
声学
计算机科学
电气工程
工程类
电信
平均曲率
流量平均曲率
几何学
谐振器
替代医学
病理
数学分析
物理
医学
数学
作者
Amirhossein Omidvar,Edmond Cretu,Robert Rohling,Mark Cresswell,Antony J. Hodgson
标识
DOI:10.1002/admt.202201316
摘要
Abstract Conventional ultrasound transducers are short and rigid, which limits their applications, especially in the area of musculoskeletal imaging where many of the structures to be imaged lie within long curved anatomical structures, such as limbs. In such cases, conformal ultrasound imaging can be advantageous. This paper presents a process for fabricating 1D and 2D flexible polymer‐based capacitive micromachined ultrasound transducers (flexible CMUTs). As a proof of concept, all elements of a 32‐element linear flexible array are shown to be functional (100% yield) and uniform in fundamental resonant frequency (SD = 1.8%). One‐way and two‐way acoustic responses during immersion tests in flat, convex, and concave array configurations (radii of curvature = 3 cm) show an average fractional bandwidth of 83% and 75%, respectively, across these bending conditions. The flexible array shows no signal drop after over 100 bending cycles and only 6% variation in pulse amplitude after over 14 h of continuous operation. Finally, the resulting transducers are shown to operate at up to 15 MHz. The findings demonstrate robust operation of flexible CMUT arrays and justify further development targeted at key imaging applications, particularly in the area of diagnosing musculoskeletal conditions.
科研通智能强力驱动
Strongly Powered by AbleSci AI