色谱法
化学
皮质酮
分析物
检出限
高效液相色谱法
萃取(化学)
固相萃取
生物化学
激素
作者
M. Hariharan,Sundar Naga,Ted VanNoord,Erick Kindt
出处
期刊:Clinical Chemistry
[Oxford University Press]
日期:1992-03-01
卷期号:38 (3): 346-352
被引量:29
标识
DOI:10.1093/clinchem/38.3.346
摘要
We have developed a simple, specific, and sensitive reversed-phase liquid-chromatographic method for accurate and simultaneous analysis of corticosterone and cortisol in human plasma. We achieved a detection limit of 300 ng/L for both steroids by modifying the old solid-phase extraction method to make use of "Tef Elutor" C18 columns, using a minibore (100 x 2 mm) analytical column, and using an ultraviolet detector with a 10-mm-pathlength flow cell. With the new extraction method absolute extraction efficiencies were greater than 90% for all the analytes, including the internal standard, flumethasone. The mobile phase was water (containing 5 mL of triethylamine per liter and citric acid to adjust the pH to 6.5), tetrahydrofuran, and acetonitrile (82/10/8 by vol). The average interassay CV for corticosterone at 0-25 micrograms/L was 6.5%; that for cortisol at 0-300 micrograms/L was 3.8%. The analytical recovery relative to the internal standard was 100.2% for cortisol and 102.6% for corticosterone. Possible interferences from drugs and other steroids were studied.