抗体
可靠性(半导体)
化学
分子生物学
色谱法
医学
免疫学
生物
功率(物理)
物理
量子力学
作者
Bert Verbruggen,I.R.O. Nováková,H. Wessels,J.B.M. Boezeman,Marijke van den Berg,Eveline P. Mauser‐Bunschoten
出处
期刊:Thrombosis and Haemostasis
[Georg Thieme Verlag KG]
日期:1995-01-01
卷期号:73 (02): 247-251
被引量:639
标识
DOI:10.1055/s-0038-1653759
摘要
Antibodies against factor VIII coagulant activity can appear in haemophiliacs who are treated with factor VIII preparations but also spontaneously in non-haemophiliacs. The Bethesda assay is the most commonly used method to detect these antibodies, but it lacks specificity especially in the lower range resulting in unreliable data. Two modifications are proposed and tested to resolve the imperfections: 1. Buffering the normal plasma used in the assay- and control mixture with 0.1 M imidazole to pH 7.4. 2. Replacing the imidazole buffer in the control mixture by immunodepleted factor VIII deficient plasma. These modifications allow better discrimination between positive and negative samples and improve reliability.