腔衰荡光谱
光谱学
吸收(声学)
一氧化氮
吸收光谱法
材料科学
氧化物
光电子学
光学
分析化学(期刊)
化学
物理
环境化学
量子力学
有机化学
冶金
作者
Lucile Richard,D. Romanini,Irène Ventrillard
出处
期刊:Sensors
[MDPI AG]
日期:2018-06-22
卷期号:18 (7): 1997-1997
被引量:28
摘要
Monitoring nitric oxide at the trace level is required in a large range of applications. We report on a trace gas analyzer optimized for nitric oxide measurements by Optical Feedback Cavity Enhanced Absorption Spectroscopy with an interband cascade laser at 5.3 µm. The short response time of the instrument allows for reaching the level of 50 ppt in only 180 ms. Its stability enables averaging up to 12 min to reach a detection limit of 0.9 ppt. Absolute concentration calibration requires to account for the optical saturation effect that results from the intense absorption line intensity addressed here, in the mid infrared region, in contrast to instruments that are operating in the near infrared region.