硬脑膜
材料科学
极限抗拉强度
生物医学工程
甘油
复合材料
医学
外科
化学
有机化学
作者
Richard van Noort,Martin Black,Terry Martin,S. Meanley
出处
期刊:Biomaterials
[Elsevier]
日期:1981-01-01
卷期号:2 (1): 41-45
被引量:98
标识
DOI:10.1016/0142-9612(81)90086-7
摘要
Heart valve replacements have now been made for some years from human dura mater. While the clinical results have been encouraging there is a considerable lack of information on the mechanical properties of the dura mater and the effects of the glycerol treatment employed. This study revealed that glycerol provided an adequate preserving medium for periods of up to 12 days, with no significant changes in the mechanical properties of dura mater. A wide range in mechanical properties was observed with, for example, the tensile strength varying from 3.28 N/mm2 to 7.86 N/mm2. Although no age-related deterioration in human dura mater could be shown, the results do suggest that dura mater in excess of fifty years of age should be avoided in order to prevent the selection of poor quality material.
科研通智能强力驱动