Lauri Koivula,Mika Kapanen,Tiina Seppälä,Juhani Collan,Jason Dowling,Peter B. Greer,Christian Jamtheim Gustafsson,Adalsteinn Gunnlaugsson,Lars Olsson,Leonard Wee,Juha Korhonen
标识
DOI:10.1016/j.radonc.2017.10.011
摘要
Background and purpose Recent studies have shown that it is possible to conduct entire radiotherapy treatment planning (RTP) workflow using only MR images. This study aims to develop a generalized intensity-based method to generate synthetic CT (sCT) images from standard T2-weighted (T2w) MR images of the pelvis. Materials and methods This study developed a generalized dual model HU conversion method to convert standard T2w MR image intensity values to synthetic HU values, separately inside and outside of atlas-segmented bone volume contour. The method was developed and evaluated with 20 and 35 prostate cancer patients, respectively. MR images with scanning sequences in clinical use were acquired with four different MR scanners of three vendors. Results For the generated synthetic CT (sCT) images of the 35 prostate patients, the mean (and maximal) HU differences in soft and bony tissue volumes were 16 ± 6 HUs (34 HUs) and −46 ± 56 HUs (181 HUs), respectively, against the true CT images. The average of the PTV mean dose difference in sCTs compared to those in true CTs was −0.6 ± 0.4% (−1.3%). Conclusions The study provides a generalized method for sCT creation from standard T2w images of the pelvis. The method produced clinically acceptable dose calculation results for all the included scanners and MR sequences.